dc.title | Evaluation and enhancement of thermal and mechanical performance of posture correction girdle for adolescent idiopathic scoliosis (AIS) | en_US |
dcterms.abstract | The posture correction girdle is a newly developed flexible orthosis to treat preteen and teenage girls between the ages of 10 and 13 who are in the early stages of scoliosis (Liu, 2014). The aim of this girdle is to enhance the physical and psychological comfort of patients in order so that there is more compliance with treatment which would increase the treatment efficacy. The girdle also enhances posture control and reduces the possibility of spinal curve progression in scoliosis. However, this functional intimate apparel still has limitations. For example, compliance is poor in the summer as the posture correction girdle is a multi-layered functional garment. If the breathability of the textile materials is inadequate, body heat and moisture cannot pass through the fabric, which would cause physiological discomfort during wear. In addition, there is little control of the spine as the girdle is very flexible, which means that the corrective forces are inadequate for the correction of spinal curvature. In the study by Liu et al. (2014) on the construction of the girdle, their subjects have a Cobb's angle that is between 6° and 20°. This means that their girdle is only suitable for those with mild scoliosis. However, there could be the need to treat scoliosis patients with more severe spinal deformities. Therefore, the girdle in Liu et al. (2014) would need to be modified to improve the thermal and mechanical properties which would extend the benefits of the treatment to those with a more severe spinal curvature. Apart from the limitations of the girdle design, there is no systematic method used to determine the most effective amount of pressure exertion and evaluate the effect of the pressure exertion on the correction of spinal deformity by the girdle. This means that Liu et al. (2014) only based their results on observation of the spinal curvature by using radiography. However, exposure to radiation may pose risks to human health. Also, the efficiency of the flexible girdle in preventing the progression of scoliotic deformity is not known after the adolescent participants wore the girdle. There is the risk that the magnitude of the spinal curvature may even increase after the wear trial. Therefore, new prediction or evaluation methods would be useful to assess the effectiveness of the modified girdle in this study. This study consists of different components, including: (a) enhancing the thermal comfort of an existing posture correction girdle in Liu et al. (2014), (b) enhancing the mechanical properties of the existing posture correction girdle, (c) a clinical study of a modified posture correction girdle with newly sourced materials, and (d) the development of prediction models including a soft mannequin with a scoliotic spine and a biomechanical model. In order to improve the thermal comfort of the existing posture correction girdle, 15 types of cool-touch fabrics in the market were newly sourced and subjected to different types of physical testing, including measuring the thermal conductivity, air and water vapor permeabilities, and peak heat flux. The thermal properties of the fabrics that were already used in the existing girdle and the newly sourced fabrics were compared to determine the most suitable combination of fabric layers. After determining the fabrics that were best suited for the modified girdle, the modified girdle was fabricated and then evaluated through the preliminary wear trial. The wear trial consisted of 2 parts - objective and subjective assessments. In the objective assessment, infrared thermography (IRT) and temperature and humidity sensors were used, while an interview was conducted for the subjective assessment. All of the results in the preliminary wear trial showed the same trend in that the modified posture correction girdle provided better thermal comfort. Therefore, the new composition of the posture correction girdle might potentially improve the thermophysiological comfort of the wearer and encourage patients to comply with the entire length of the treatment. In order to enhance the mechanical properties of the existing posture correction girdle, 4 different types of supporting struts in the market were newly sourced and along with the struts used in the original girdle in Liu et al. (2014), subjected to bending and recovery tests to compare their mechanical properties. The most suitable material that would improve the efficacy of spinal correction treatment with a girdle should be adequately stiff with a good recovery property. A preliminary wear trial which included pressure measurements and radiographic examination was also conducted to determine the effect of the girdle with different types of struts. Based on the findings, modified versions of the girdle with moulded shape memory alloy (SMA) struts were developed since the SMA struts had the best recovery and a high degree of stiffness. However, they could not exert pressure evenly because they did not easily bend to conform to the curves of the body. For orthoses that require an even pressure distribution, moulding in accordance with the body shape was recommended. | en_US |
dcterms.abstract | A clinical study was also carried out to further study the efficacy of the modified posture correction girdle. In total, 16 female subjects with adolescent idiopathic scoliosis (AIS) participated in this study, among which 6 wore the original posture correction girdle while 10 wore the modified girdle. Among the 10 subjects in the group who wore the modified girdle, 5 of them wore the girdle with curved SMA struts on one side and the other 5 wore the girdle with curved SMA struts on both sides. The results showed that the insertion method with curved SMA struts inserted into both sides of the girdle provided the best corrective effect while the girdle with curved SMA struts on one side had the poorest corrective effect among the three experimental groups. The result proved that the hard brace mechanism which was adopted in the girdle with curved SMA struts on one side was not feasible. On the other hand, the perceived comfort of the girdles was assessed by conducting a questionnaire and interviews. The results indicated that the girdle inserted with curved SMA struts on one side was problematic in terms of psychological and physiological comfort. This girdle was unbalanced at the waist and was a concern because the wearers might be the object of scrutiny and curiosity due to their appearance. Therefore, the girdle might increase the stress level of the subjects. Also, straight SMA struts caused pain and pose difficulties of stretching and fixing the waistband into the correct position. Therefore, the perceived comfort of the girdle inserted with curved SMA struts on one side was the lowest while the girdle inserted with resin bone (RB) struts was the highest in perceived comfort due to the high flexibility, lightness in weight and minimal restriction of body movement. After obtaining the clinical data, prediction models were developed to simulate the treatment with a girdle on a scoliotic patient. In order to validate the biomechanical model, a soft mannequin was first developed and validated. The soft mannequin was a moulded silicone rubber torso which contained three-dimensional (3D) printed vertebrae and intervertebral discs. These components were the same as those in the finite element model (FEM) formulated in this study. Two prototypes were constructed. The validation of the soft mannequin was done by pressure measurements and spinal curve changes which were compared to the clinical results. The results showed that the soft mannequin made of softer silicone rubber (Shore 00 hardness of 10) produced similar pressure results as those of the real-life cases. Similar to the clinical study, there were greater shifts in the spine when the girdle with SMA struts was worn, so that it could be concluded that the corrective effect of the girdle with SMA struts excelled that of the girdle with RB struts. With regard to the FEM, two models were constructed to simulate the posture correction girdle with SMA and RB struts. Both models consisted of five geometric models: the torso, posture correction girdle, supportive struts, vertebrae of the thoracolumbar/lumbar spine and intervertebral discs. The FEM was validated by using the interface pressure and spinal corrective effects measured from the validated soft mannequin model. The output of the FEM showed extremely low stress values compared to the measured pressure of the soft mannequin. Although the pressure values obtained from the FEM and soft mannequin differ at the pelvis, they showed the same trend in that the interface pressure on the left side of the underbust was higher than that at the right side of the underbust, the interface pressure at the right side of the waist was higher than that at the left side of the waist and the left side of the pelvis was higher than that of the right side of the pelvis for the girdle with RBs. However, in terms of the spinal corrective effects, the FEM results were not comparable with those of the soft mannequin model. The model contains many sharp edges and small meshes which might affect the calculations. The research results and findings provided useful information on the effects including the thermal and mechanical performances since the flexible scoliotic brace (girdle) is constructed of different types of materials. In addition, the relationship between material stiffness and interface pressure was useful for future developments in flexible orthoses. The modified version of the posture correction girdle showed better thermal comfort and corrective effect than the original design. Although the efficacy of the modified girdle was not comparable with conventional hard brace treatment, this study provided a reference for improving the flexible brace treatment. The soft mannequin could also be used to predict the pressure distribution and corrective effect of the posture correction girdle. Finally, the outputs of this project could extend to the development of a flexible brace that provided wear comfort and good corrective effects on scoliotic patients. | en_US |
dcterms.extent | xxxi, 287 pages : color illustrations | en_US |
dcterms.isPartOf | PolyU Electronic Theses | en_US |
dcterms.issued | 2019 | en_US |
dcterms.educationalLevel | Ph.D. | en_US |
dcterms.educationalLevel | All Doctorate | en_US |
dcterms.LCSH | Hong Kong Polytechnic University -- Dissertations | en_US |
dcterms.LCSH | Scoliosis -- Patients -- Treatment | en_US |
dcterms.LCSH | Scoliosis in children | en_US |
dcterms.LCSH | Orthopedic braces | en_US |
dcterms.accessRights | open access | en_US |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
991022289511703411.pdf | For All Users | 8.79 MB | Adobe PDF | View/Open |
Copyright Undertaking
As a bona fide Library user, I declare that:
- I will abide by the rules and legal ordinances governing copyright regarding the use of the Database.
- I will use the Database for the purpose of my research or private study only and not for circulation or further reproduction or any other purpose.
- I agree to indemnify and hold the University harmless from and against any loss, damage, cost, liability or expenses arising from copyright infringement or unauthorized usage.
By downloading any item(s) listed above, you acknowledge that you have read and understood the copyright undertaking as stated above, and agree to be bound by all of its terms.
Please use this identifier to cite or link to this item:
https://theses.lib.polyu.edu.hk/handle/200/10231